Products
Pollutant Source VOCs Monitor
Pollutant Source VOCs Monitor
MTHC/nMTHC Analyzer PUREAIR 50

MTHC/nMTHC Analyzer PUREAIR 50

PUREAIR 50 MTHC/nMTHC analyzer, adopts gas chromatography FID(GC-FID) detection method. It is of high sensitivity and stable performance, mainly used for online monitoring of MTHC/nMTHC in pollution sources and exhaust emissions.

  1. Low detection limit up to 0.1ppm.
  2. Stable performance and good RSD.
  3. Flexible application with versatile solutions.
  4. Requires small amount of sample(1ml),and less consumables.

Technical Parameter

  1. Detection method: FID
  2. Detection objectives: MTHC, nMTHC
  3. Measuring range:0~50/500/1000/10000ppm (programmable)
  4. Low detection limit: 0.1ppm
  5. Repeatability standard deviation(RSD): ≤2.0%
  6. Span drift: ±2.0% FS
  7. Analysis cycle: ≤5 min

Online BTEX Analyzer  PUREAIR 60

Online BTEX Analyzer PUREAIR 60

PUREAIR 60 online BTEX analyzer, adopts gas chromatography FID(GC-FID) detection method. It is of high sensitivity and stable performance, mainly used for online monitoring of toluene, ethylbenzene, xylene, styrene in pollution source and exhaust emissions.

  1. Low detection limit up to 0.1ppm.
  2. Stable performance and good RSD.
  3. Flexible application with versatile solutions.
  4. Require small amount of sample (1ml), and less consumables.

Technical Parameter

  1. Detection method: FID
  2. Detection objective: benzene, toluene, xylene
  3. Measuring range: 0~20/50/500/1000ppm (programmable)
  4. Low detection limit: 0.01ppm
  5. Repeatability standard deviation (RSD): ≤2%
  6. Span drift: ±2.0% FS.
  7. Analysis cycle:
    ≤15min (Quantitative loop injection)
    ≤60min(Adsorption tube injection)